瀝青閃點(diǎn)儀英文版說(shuō)明書

First, the applicable standard

This instrument was petroleum and petroleum products test method standards T3536-2000 GB/ "petroleum products-determination of Flash and fire (Cleveland Open Cup method) to design and manufacture.

This instrument also applies to JTJ052 2000 highway code T0611 2000 the asphalt for Flash and fire test (Cleveland Open Cup method) and the international prospective ISTMD 2000 test.

2. purpose and scope of application

The tester is applied to Cleveland Open Cup (hereinafter referred to as COC) determination of viscous oil bitumen, coal tar pitch and test 79 ° c above the liquid asphalt material for Flash and fire safety for construction use.

III. instruments and working conditions

        Power supply: 220V ± 10% 50H

        Ambient temperature: -10 ¡æ ~ 50 ¡æ

        Relative humidity: 85%

Fourth, the main features and technical parameter

 Automatic sweep stroke, ignition;

 Electric heating power 0 ~ 400W continuously adjustable;

 Heating wire consists of transparent quartz tube protection, no open flames, explosions, and the temperature rises rapidly;

5. block diagram of the main structure and principle

 

Structure, as Figure 1

                  Figure 1

Principle diagram of figure 2

Figure 2

VI. installation and use

1 , Open box, press the instrument integration and technical documentation to check instrument appearance and related accessories, such as damage or missing, according to the chart shown in the installation.

2 , Check the power supply and grounding is good.

3 , Turn the power switch, the led lights, then you can carry out test operations. 4, specimen Cup with solvent cleaning, drying, installed in the bracket.

5 , Installing thermometers, thermometers, vertical specimen Cup is inserted, the thermometer mercury bulb from the Cup about 6.5mm, location and relative side margin igniter glass edge 16mm.

6 , Sample preparation, sample Cup to marking and other parts of the sample Cup not stained with asphalt. (Note: the sample heating temperature should not exceed the following 55 flash point ℃)

7 , All equipment should be placed in the Interior light darker and no significant air circulation, and three-sided enclosure windscreen.

8 , Ignitor and gas-pressure vessel (e.g. liquefied propane gas or natural gas) with hose connection, turned the ignition, flame ignition testing, debugging into standard ball shape or diameter 4mm ± 0.8mm small spherical flame test.

7. test procedure

1 , Start heating test, so that the heating rate rapidly achieve 14 ¡æ /min ~ 17 ¡æ/min. When the sample temperature is expected to achieve before 56 flash point ℃, adjust the heater to reduce heating rate so that the expected Flash before 28 ° c can make heating rate control in 5.5 ¡æ /min ± 0.5 ¡æ /min.

2 , Sample temperature desired Flash before 28 ℃, every 2 ° c according to the ignition button once so that the flame igniter test centers along the test Cup 150mm RADIUS for arc level scan once, from the experimental side of the Cup to the other side of the elapsed time of about 1s. You should confirm that the pilot flame Igniter for diameter 4mm ± 0.8mm fireball, and is located in the test above 2.5mm Cup. (Note: the test is not a specimen Cup)

3 , When initially appears on the specimen surface in a moment that is off the blue flame, immediay credited from thermometers, temperature on reading as specimens of Flash. Note do not try to flame around the blue and white flame mistaken for the Flash fire.

4 , Continue to maintain sample heating, heating rate 5.5 ¡æ /min ± 0. 5 ¡æ /min, and press the required test CDI ignition.

5 , Immediay when the sample contact flame ignition and combustion can continue for a period of not less than 5s, stop heating, and read the temperature on the recording thermometer as a specimen.

6 At least the same specimen , parallel test twice, two measuring results is less than the margin of a repetitive test allows poor 8 ℃, the average value of the integer as a test result.

7 , Precision or allow poor

Repetitive test allows the poor to: flash point 8 ℃, lighted 8 ℃;

Reproducibility test allows for: flash point 16 ℃, lighted 14 ℃.

8 , After the end of the test, as well as the cleaning and should cut off the power.

版權(quán)所有© 2018 滄州路儀試驗(yàn)儀器有限公司

冀公網(wǎng)安備 13092902000113號(hào)

日韩欧美成人免费观看| 欧美日韩一区二区三区在线| 亚洲综合另类小说色区| 亚洲综合另类小说色区| 国模无码一区二区三区不卡| 久久夜色撩人精品国产| 西西4444WWW大胆无码| 欧美日韩一区二区三区在线| 在线观看免费av网| 日韩三级黄片| 精品国产欧美一区二区三区成人| 黑人巨粗进入警花肖雨霖| 欧洲精品乱码久久久久久| 亚州少妇| www.色日本| 麻阳| 国产三级视频在线播放线观看| 无套内谢的新婚少妇国语播放| 无码一区二区三区老色鬼| 无套中出丰满人妻无码| 亚洲熟妇丰满多毛XXXX| 午夜色大片在线观看| 国产成人综合在线视频| 亚洲色偷偷综合亚洲AVYP| 亚洲最大激情中文字幕| 欧美第十页| 97激情| 99热精品在线观看| 隆尧县| 鲁丝片一区二区三区免费| 亚洲精品国产精品国自产观看| 狠狠噜天天噜日日噜无码| 精品国产精品国产偷麻豆| 免费1级欧美黑寡妇a观看| 狠狠色丁香久久婷婷综合| 91亚洲一线产区二线产区| 你懂的网址在线| xxx窝窝美女色www| 国产午夜成人久久无码一区二区| 欧美日韩国产一区二区三区不卡 | 亚洲日韩aⅴ在线视频|